Content of Premarket Submissions for Device Software Functions
2021-11-01
FDA-2021-D-0775
其他
美国
征求意见稿或草案
2021-11-01
FDA
This guidance document is intended to provide information regarding the recommended documentation sponsors should include in premarket submissions for FDA's evaluation of the safety and effectiveness of device software functions, which are functions that meet the definition of a device under section 201(h) of the Federal Food, Drug, and Cosmetic Act (FD&C Act).1 The recommendations in this guidance document pertain to device software functions, including software in a medical device (SiMD) and software as a medical device (SaMD).2 When final, this document will replace FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices issued on May 11, 2005, and it will update FDA's thinking related to the documentation FDA recommends sponsors include for the review of device software functions in premarket submissions.
This guidance identifies the software information generally necessary for evaluating the safety and effectiveness of a device in a premarket submission. The recommendations in this guidance also may help facilitate FDA's premarket review. This guidance describes information that would be typically generated and documented3 during software development, verification, and design validation. The least burdensome approach was applied to identify the minimum amount of information that, based on our experience, would generally be needed to support a premarket submission for a device that uses software. During premarket review, FDA may request additional information that is needed to evaluate the submission. For example, in order to demonstrate a reasonable assurance of safety and effectiveness for devices that use software, documentation related to the requirements of the Quality System Regulation (QSR) (21 CFR Part 820) is often a necessary part of the premarket submission. As part of QSR design controls, a manufacturer must "establish and maintain procedures for validating the devices design," which "shall include software validation and risk analysis, where appropriate." (21 CFR 820.30(g)).
The documentation recommended in this guidance is based on FDA's experience evaluating the safety and effectiveness of device software. However, sponsors may use alternative approaches and provide different documentation so long as their approach and documentation satisfies premarket submission requirements in applicable statutory provisions and regulations. For the current edition of the FDA-recognized consensus standard(s) referenced in this document, see the FDA Recognized Consensus Standards Database. For more information regarding use of consensus standards in regulatory submissions, please refer to the FDA guidance titled Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices and Standards Development and the Use of Standards in Regulatory Submissions Reviewed in the Center for Biologics Evaluation and Research.
1 The term "device" is defined in 201(h) of the Federal Food, Drug, and Cosmetic (FD&C) Act to include an "instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other similar or related article, including any component, part, or accessory, which is ...intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, in man ... or intended to affect the structure or any function of the body of man..." and "does not include software functions excluded pursuant to section 520(o) of the FD&C Act."
2 See FDA website on "Software as a Medical Device (SaMD)."
3 As a reminder, manufacturers of device software must create and maintain software-related documentation in accordance with the requirements of the Quality System (QS) Regulation (21 CFR 820.30 Subpart C – Design Controls of the Quality System Regulation).
最新政策法规资讯
-
摩熵医药2026-06-17
-
摩熵医药2026-06-09
-
赛柏蓝2026-06-01
-
深蓝观2026-05-22
-
赛柏蓝2026-04-24
-
摩熵医械2026-01-28
-
摩熵医械2026-01-13
-
摩熵医械2025-12-19
-
赛柏蓝2025-12-12
-
摩熵医械2025-11-28
热门政策标签
医药市场洞察中心
医药市场销售数据分析 · 智能洞察 · 竞争格局监测
-
医药市场销售数据分析 · 医院、实体药店、网上药店
-
市场竞争格局分析(靶点、药品类型、ATC大类)
-
2026-08-31
-
2026-08-31
-
摩熵咨询 35 45
-
摩熵咨询 28 46
-
颁布机构:国家市场监督管理总局 2026-08-10
-
颁布机构:国家市场监督管理总局 2026-05-29
最新其他政策
-
关于做好2026年基本公共卫生服务工作的通知
2026-09-01
-
放射性粒籽植入防护产品注册审查指导原则(征求意见稿)等55项第二类无源医疗器械注册审查指导原则意见
2026-09-01
-
负压源压力源吸引器注册审查指导原则(征求意见稿)等59项第二类有源医疗器械注册审查指导原则意见
2026-09-01
-
甲状旁腺激素测定试剂注册审查指导原则(征求意见稿)等16项第二类体外诊断试剂医疗器械注册审查指导原则意见
2026-09-01
-
单克隆抗体动物试验减少或替代指导原则(征求意见稿)
2026-08-31
-
黑龙江省省本级门诊慢性病病种认定标准(征求意见稿)
2026-08-31
-
关于全国中成药联采部分中选药品信息变更的通知
2026-08-31
-
天津市民政事业发展“十五五”规划
2026-08-31
-
关于执行第四批和第二批接续全国中成药采购联盟集中带量采购中选结果的通知
2026-08-31
-
关于公开征求《全身型重症肌无力治疗药物临床试验技术指导原则(征求意见稿)》意见的通知
2026-08-28
-
关于做好全国第四批中成药采购联盟等批次中选结果执行工作的通知
2026-08-27
-
做好国家组织冠脉支架集中带量采购第二轮接续采购结果执行工作
2026-08-27
-
进一步优化药品和医用耗材集中带量采购管理工作
2026-08-26
-
关于做好长期护理保险支付管理工作的指导意见
2026-08-26
-
关于规范整合口腔等4类医疗服务价格项目的通知
2026-08-26
全球药物研发中心
全球研发数据与行业前沿情报
-
2026-08-31
-
2026-08-31
-
2026-08-31
-
摩熵咨询 20 8
-
摩熵咨询 34 36
-
摩熵咨询 20 5
-
颁布机构:四川省经济和信息化厅、中共四川省委军民融合发展委员会办公室、四川省教育厅、四川省科学技术厅、四川省人力资源和社会保障厅、四川省卫生健康委员会、四川省医疗保障局、四川省中医药管理局、四川省药品监督管理局 2026-08-04
-
颁布机构:国家医保局 2026-07-31
-
颁布机构:国家药品审评中心(CDE) 2026-07-24